Spikes aren’t showing up
- Verify that study is processed
- Click on Track Acquisition and compare time shown in trends to time at end of study
- Manually check for trend results files in study location
- Change sensitivity to High
- This may indicate whether anything was marked at lower perception values
- Verify any issues with the recording workflow
- Verify that study was recorded using 10-20 channel labels
- Verify that study does not have a large amount of open channel EEG or calibration in first 5 minutes of recording
- Manually identify something that should have been marked and contact support@persyst.com.